Simco Fmx002 Manual Top !!top!! -
To maximize the performance and efficiency of your Simco FMX002 Manual Top:
) indicates a negative charge; no sign indicates a positive charge. If the measurement exceeds ±plus or minus 20kV, the screen will display "Err" or flash. 5. Tips for Maximum Accuracy (Top Tips) simco fmx002 manual top
Verify that the FMX-002 reads between +0.90 kV and +1.10 kV (accounting for the ±10% accuracy tolerance). To maximize the performance and efficiency of your
: Features a unique dual LED guide ring light system. When the two red spots overlap on the target surface, it confirms the device is at the correct 1-inch (25mm) measuring distance. Tips for Maximum Accuracy (Top Tips) Verify that
If heavy contamination occurs, the unit must be sent to an authorized SIMCO service facility for cleanroom teardown and cleaning. Annual Calibration Verification
overlap to form a single concentric circle. This indicates the standard 1-inch (25mm) measuring distance. : 0 to ±1.49 kV. High Range : ±1.0 kV to ±20.0 kV.
Flashes or stays illuminated when the 9V battery needs replacement. Standard Operating Procedure